The Elara 11 is designed for sterilisation of wrapped, porous and hollow loads. The chamber volume is large, external dimensions are small, making it especially suitable for clinics requiring a high turnover. The control systems are easy to use.
For consistent and reliable performance choose Tuttnauer.
Efficient
- Short cycle and high volume pre and post vacuum pump for perfect class B sterilisation results.
- 316L grade stainless steel chamber and door with electro-polish finish
- A built-in steam generator provides steam and facilitates a standby steam mode
- The door closes smoothly and locks automatically
Features
- Large 28.5 Liter chamber within a small foot print. A perfect fit for standard 60 cm shelves or counters
- Provided with 5 stainless steel trays and can also hold cassettes
- Water filling from the front economises on bench space by eliminating the need to keep the top of the autoclave accessible
Safety Features
- A cycle will not start unless the door is properly locked
- An automated double locking safety device prevents the door from opening while the chamber is pressurized
- An automatic safety shut-off prevents overheating of the chamber
- The autoclave comes with safety pressure-release valves for the chamber and generator
Technical Specifications |
|
Chamber Dimensions |
280 x 504 |
Chamber Volume L |
28.5L |
Number of Trays |
5 |
Tray Dimensions (WxHxD) |
194 x 22 x 442 |
Voltage(V) Freq.(Hz) |
1ph 230 V 50/60 Hz |
Current(A) Power(W) |
10 A 2300 W |
Overall Dimensions (WxHxD) |
530 x 444 x 645 |
Autoclave Weight (Kg.) |
66 |
Sterilisation Cycles
The Elara 11 offers a total of 8 programs: 6 sterilisation programs and 2 test programs: Bowie and Dick Test and Vacuum Test.
Program |
Instrument and Materials |
Sterilization Temp |
Sterilization Time |
Drying Time |
Average Cycle Time |
Flash 134 |
Unwrapped Instruments |
134°C (273°F) |
4 Min. |
1 Min. |
24 Min. |
P 134 |
Porous Wrapped and Hollow A |
134°C (273°F) |
18 Min. |
15 Min. |
52 Min. |
Wdry 134 |
Wrapped Instruments |
134°C (273°F) |
4 Min. |
15 Min. |
37 Min. |
NoDry 121 |
Unwrapped Instruments |
121°C (250°F) |
20 Min. |
- |
38 Min. |
WDry 121 |
Porous, Wrapped and Hollow A |
121°C (250°F) |
20 Min. |
20 Min. |
58 Min. |
Delicate |
Porous, Wrapped and Hollow A |
121°C (250°F) |
20 Min. |
- |
35 Min. |
The Elara 11 steriliser complies with the strictest international standards and directives:
Technical Directives
- Medical Device Directive 93/42/EEC
- Pressure Equipment Directive 97/23/EEC
- WEEE and RoHS
Technical Standards
- EN 13060:2004
- EN 61010-1
- EN 61010-2-041
- EN 61326
- EN 61010-2-040
- ISO 17665
Quality Standards
The manufacturing plant meets the following quality standards:
- ISO 9001:2000 – Quality Management Systems
- ISO 13485:2003 – Medical devices
Related information: